EasyManua.ls Logo

Honeywell BW GasAlertMicro 5 - Replacing the Pump Filter (Generation 2 Pump)

Honeywell BW GasAlertMicro 5
113 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
67
GasAlertMicro 5/PID/IR
Installing the Pump Module
Replacing the Pump Filter (Generation 2 Pump)
To replace the Generation 2 pump filter, refer to Figure 7. and the follow-
ing steps 1-8.
Note
If using the auxiliary filter on the Generation 1 pump or as an
additional filtration system on the new pump module, refer to
Attaching the Auxiliary Filter
.
Figure 7. Replacing the Pump Filter (Generation 2 Pump)
1. Deactivate the detector.
2. Remove the filter window screw.
3. Remove the filter window.
4. Remove the old filter from the filter window.
Important! Ensure the filter cavity and filter window are clean
and free of debris.
5. Insert a new filter.
6. Replace the filter window and screw.
7. Activate the detector. The detector performs a pump test
during the startup self-tests to verify the pump module is oper-
ating correctly.
8. To order additional filters, refer to Replacement Parts and
Accessories.
a Caution
Filters may need to be replaced more frequently in high
particulate areas. BW Technologies by Honeywell
recommends that the auxiliary filter be used as a
additional filtration in high particulate areas.
www.GlobalTestSupply.com