GasAlertMicro 5/PID/IR
User Manual
66
Installing the Pump Module
Figure 6. Installing the Pump Module
Table 14. Installing the Pump Module
To install the pump module, refer to Figure 6., Table 14., and the
following procedures.
1. Deactivate the detector.
2. Remove the two machine screws and the sensor cover.
Remove the sensor filter from the sensor cover and insert it
into the pump module.
3. Attach the pump module to the detector and replace the two
machine screws.
4. If sampling in a confined space, attach the sample probe. Refer
to Confined Space Sampling
.
If using an auxiliary filter as an additional filtration system, refer
to Attaching the Auxiliary Filter
.
5. Activate the detector. The detector performs a pump test
during the startup self-tests. Refer to Pump Test
.
Item Description
1 Pump module
2 Sensor filter
3Detector
4 Machine screws (2)