GasAlertMicro 5/PID/IR
User Manual
60
Setting the Low Alarm Setpoint
The current low alarm setpoint displays for the selected sensor.
Press H or G to change the value for the low alarm setpoint. When the
required value displays, press C to confirm.
Setting the High Alarm Setpoint
The current high alarm setpoint displays for the selected sensor.
Press H or G to change the value for the high alarm setpoint. When the
required value displays, press C to confirm.
Proceed to step #9 Setting the Remaining Alarm Setpoints
.
Setting the Remaining Alarm Setpoints
9. Repeat step #8 to set the alarm setpoints for the remaining
sensors. When complete, the detector emits two quick beeps
and proceeds to Finish Calibration
.
Finish Calibration
10. The detector displays the following screen to indicate that
calibration is complete.