EasyManua.ls Logo

Honeywell BW MaxXT II - Calibration Requirements; Calibration Gas Cylinders and Regulators; Calibration Diagnostics Protection

Honeywell BW MaxXT II
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER
6
Calibration
Guidelines
Recommended gas mixture:
CO: 100 ppm balance N
2
H
2
S: 25 ppm balance N
2
LEL: 50% LEL or 2.5% by vol. methane balance air
O
2
: clean air, 18%
l To ensure accurate calibration, use a premium-grade calibration gas. Gases approved by
the National Institute of Standards and Technology (NIST) improve the validity of the
calibration. You can use a 34 L cylinder (PN: CG-Q34-4) or a 58 L cylinder (PN: CG-Q58-4)
l Do not use a gas cylinder past its expiration date.
l Calibrate a new sensor before use. Install the sensor, activate the detector, and allow the
sensor to stabilize before starting calibration or a bump test.
Used sensor: 60 seconds / new sensor: 5 minutes.
l Calibrate the detector at least once every 180 days, depending upon use and sensor
exposure to poisons and contaminants.
l Calibrate the detector if the ambient gas varies during startup.
l Calibrate only in a safe area that is free of hazardous gas.
l Do not calibrate the detector during or immediately after charging is complete.
l The H
2
S (PN: SR-H-MC), CO (PN: SR-M-MC), LEL (PN: unfiltered SR-W-MP75; filtered SR-
W-MP75C), and O
2
(PN: SR-X2V), sensors can be automatically zeroed each time the
detector is activated if the Auto-Zero on Startup option is enabled. Activate the detector in a
normal (20.9%/20.8% O
2
) atmosphere.
l If a certified calibration is required, contact BW Technologies by Honeywell.
l Calibration can be performed using either a 0.5 l/min manual regulator (PN: REG - 0.5) or
a demand flow regulator (PN: REG-DF-1)
l The maximum hose length for calibration is 3 ft (0.9 m).
BW MaxXT II 70 User Manual

Other manuals for Honeywell BW MaxXT II

Related product manuals