EasyManuals Logo

Honeywell BW MaxXT II User Manual

Honeywell BW MaxXT II
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #78 background imageLoading...
Page #78 background image
l the gas concentration drops below 50% of the expected gas level during the 2-
minute span.
If gas is applied to a sensor and the detector fails to span the sensor, perform the
following:
l Repeat the calibration using a new gas cylinder.
l Repeat the calibration using a new 0.5 l/min regulator or a new demand flow
regulator.
l If the sensor fails the span a second time, replace the sensor. Refer to See
"Replacing a Sensor or Sensor Filter" on page91 for more information..
If a sensor(s) fails any step of the calibration, the following screen displays.
Note: If calibration is unsuccessful for a sensor(s), the calibration due date cannot be
set for the failed sensor(s).
BW MaxXT II 77 User Manual

Other manuals for Honeywell BW MaxXT II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell BW MaxXT II and is the answer not in the manual?

Honeywell BW MaxXT II Specifications

General IconGeneral
BrandHoneywell
ModelBW MaxXT II
CategoryGas Detectors
LanguageEnglish

Related product manuals