7. Enter the span gas concentration values. The values entered in the Calibrate Device
popup must match the span concentration values on the gas cylinder.
8. Click inside the checkbox for each sensor that will be calibrated, and then click Start
Test.
9. Refer to See "Auto Span" on page76 for more information. to complete calibration.
Note: To calibrate using the IntelliDoX, refer to the IntelliDoX User Manual and the Safety Suite
Device Configurator (SSDC) Operator’s Manual.
BW MaxXT II 81 User Manual