Getting Started
Part No.: 4417584_Rev12 Configuration Manual
CIU 888 3 - 58
$*6
ASTM-IP-52 T23/24, ASTM-IP-52 T53/54
API MPMS Ch. 11.2.4 (2007) (GPA TP-27-07) T23/24, API MPMS
Ch. 11.2.4 (2007) (GPA TP-27-07) T53/54, API MPMS Ch. 11.2.4
(2007) (GPA TP-27-07) T59/60
ASTM D1555-2016 API @ 60 F, ASTM D1555M-2016 kg/m3 @ 15 C
EN15940-2019, SGS-2021, NBR15639-2016
c) For NBR15639-2016, the order of precedence is Reference
density, mass concentration and Sample density & sample
temperature. Set the Density input to N.A, to set mass concen-
tration as the input.
d) For SGS-2021, the order of precedence is Reference density,
Sample density & sample temperature.
e) For EN15940-2019, the order of precedence is Reference density,
Sample density & sample temperature.
f) For User Configurable Density table, associate a user defined
density table by right clicking product site element and selecting
“Add density table” option. Refer to Appendix E for the format of
the user defined density table.
g) For User Configurable CTL table, associate a user defined density
table by right clicking product site element and selecting “Add CTL
table” option. Refer to Appendix F for the format of the user
defined CTL table.
h) For User Configurable Concentration table, associate a user
defined density table by right clicking product site element and
selecting “Add concentration table” option. Refer to Appendix G
for the format of the user defined concentration table.
7. Click the Product Operational Settings tab, the Properties detail
appears.