3. Installation
3.5. Conduit / Cable Entries
14 SmartLine Wireless Temperature and Universal I/O User's Manual Revision 1
3.5 Conduit / Cable Entries
NOTICE
THIS PRODUCT IS SUPPLIED WITH PLASTIC DUST PLUGS IN THE
CONDUIT/CABLE GLAND ENTRIES. IT IS THE USERS
RESPONSIBILITY TO PROVIDE CABLE GLANDS, ADAPTORS AND/OR
BLANKING PLUGS SUITABLE FOR THE ENVIRONMENT IN WHICH THIS
PRODUCT IS INSTALLED. WHEN INSTALLED IN A HAZARDOUS
LOCATION THE CABLE GLANDS, ADAPTORS AND/OR BLANKING
PLUGS SHALL ADDITIONALLY BE SUITABLE FOR THE HAZARDOUS
LOCATION, THE PRODUCT CERTIFICATION AND ACCEPTABLE TO
THE AUTHORITY HAVING JURISDICTION FOR THE INSTALLATION
Summary
Table 3-3: Conduit entry plugs and cable glands for your transmitter.
Factory Part
No.
Description
Environmental
rating
Ambient
Hazardous Location
Certification
50000547-001
M20 Conduit
Plug
IP66-68, 4X,
6P
–40 – 85ºC
–40 – 185ºF
ATEX EEx de IIC
50021832-002
½ NPT Conduit
Plug
IP66-68, 4/4X,
6/6P
–40 – 85ºC
–40 – 185ºF
ATEX EEx d IIC;
CSAcus CL I, Zone 1, Ex/AEx d
IIC; CL I, Div 1 & 2, GP ABCD;
CL II, Div 1 & 2, GP EFG; CL III,
Div 1 & 2
50023232-001
M20 Cable
Gland
IP68
–40 – 100ºC
–40 – 212ºF
-----------------------------------
50023212-001
½ NPT Cable
Gland
IP68
–40 – 100ºC
–40 – 212ºF
-----------------------------------
Note: Torque conduit entry plugs, cable glands, or adapters to 32 Nm (24 lb-ft).