1- 14 430-95597-001 (Rev. 06/05)
Maintaining a minimum separation distance of 2 ½ ft
between the Infusion pump system and portable/
mobile RF communications equipment
List Numbers 12391 and 11971 are compliant to IEC/EN
60601-1-2 (1993)
List Numbers 20678 & 20679 are compliant to IEC/EN
60601-1-2 (2001) and have been tested and found to comply
with EMC limits for the Medical Device Directive 93/42/EEC
(EN 55011 Class B and EN 60601-1-2:2001).
Guidance on EMC Compatibility
There is a shared responsibility between manufacturers,
customers and users to ensure that Medical Equipment and
Systems are designed and operated as intended. Medical
electrical equipment needs special precautions regarding
electromagnetic compatibility and needs to be installed and
used according to the electromagnetic compatibility
information provided in this manual.
The device is suitable for use in all establishments, including
domestic establishments. If extended operation during
power mains interruption is needed, use battery power.
Always manage the electromagnetic environment.
The guidance included in this manual provides information
needed to:
Determine the device’s suitability for use in the
intended environment.
Manage the electromagnetic environment to permit the
device to perform as intended without disturbing other
equipment.
Separate the device from all other electronic equipment. If
the device must be used near other electrical equipment,
monitor the equipment to ensure there is no electromagnetic
interference.
Devices should not be used adjacent to or stacked with other
equipment. If the device must be used adjacent to or stacked