3- 10 430-95597-001 (Rev. 06/05)
4 Push in the flow regulator to close. Confirm
there is no flow.
Loading the cassette
To load the primed cassette into the plum A+, proceed as
follows.
1 Open the cassette door by lifting up the
handle.
2 Insert cassette into door guides. Close the
cassette door. Confirm there is no flow and
no kinks appear in tubing.
NOTE: If flow is observed, close the tubing
clamps and replace the administration set.
NOTE: The administration set should be
changed per CDC guidelines or healthcare
provider policy. Discard after use.
Preparing the Secondary Line
CAUTION: CONSULT THE DRUG LABELING TO CONFIRM DRUG
COMPATIBILITY
, CONCENTRATION, DELIVERY RATES, AND VOLUMES
ARE
ALL SUITABLE FOR CONCURRENT DELIVERY OR PIGGYBACK
sec03.fm Page 10 Wednesday, June 8, 2005 8:19 AM