4- 2 430-95597-001 (Rev. 06/05)
BE DISRUPTED. OPERATION OF THE INFUSER UNDER THESE
CONDITIONS
SHOULD BE AVOIDED.
TANDEM CARRIER INSTRUCTIONS
NOTE: The Plum A+3 is NOT to be used with the tandem
carrier. It is only to be used with the Plum A+ and other single-
channel devices.
The Plum A+ may be safely and conveniently mounted on an IV
stand or on a tandem carrier mounted on an IV stand.
Always mount the tandem carrier
(List# 12270-01) to the pole before
attaching the infuser(s) to the tandem
carrier.
Make sure tandem carrier clamp is
attached firmly to pole.
Make sure infusers are firmly
attached to tandem carrier.
This section details the Plum A+
instrument setup procedures.
SYSTEM SELF-TESTS
The infuser performs a suite of System Self-Tests when power
is applied to confirm readiness for use.
Failure during the Self-Tests is reported as a Malfunction
Condition.
CAUTION: D
O NOT PLACE THE PLUM A+ IN SERVICE IF IT FAILS
THE
SYSTEM SELF-TESTS.
Connect the AC power (mains) cord to an AC power receptacle,
then confirm that the Charging/Line indicator illuminates (above
the [ON/OFF] key on the front panel).
A systematic self-testing of the processing, delivery, and safety
systems is performed whenever the Plum A+ is turned on, to
verify readiness for operation.
sec04.fm Page 2 Wednesday, June 8, 2005 8:22 AM