For Systems With Hospira MedNet
TM
Software
4- 20 430-95597-001 (Rev. 06/05)
3 To resume infusion,
press .
4 Select
[Start A],
[Start B], or
[Start All].
Or, to discontinue infusion,
press .
5 Clamp secondary tubing if connected.
6 Open door, remove set and close door.
Backpriming
BACKPRIMING- Removal of air or
fluid from the proximal line and
cassette airtrap.
1 To initiate a backprime, select
[Back Prime].
NOTE: Backpriming is available only
when all delivery is stopped.
Secondary container is necessary to
receive the air.
AB
STOPPED STOPPED
Rate
mL/hr
Vol Inf
mL
Cancel
25
3 20.
4
50
Start
All
Start
A
Start
B
Choose line(s) to start:
! !
MedSurg
A
A
B
B
STOPPED STOPPED
Rate
mL/hr
Vol Inf
mL
Options/
Vol Inf
5.5
8.
9
0
0
Back
Prime
!
MedSurg
sec04.fm Page 20 Wednesday, June 8, 2005 8:22 AM