Plum A+
®
Infusion Systems
430-95597-001 (Rev. 06/05) 4- 15
Concurrent Delivery
CONCURRENT DELIVERY-
Simultaneous delivery of two fluids at
independent flow rates.
CAUTION: C
ONSULT DRUG
LABELING
TO CONFIRM DRUG
COMPATIBILITY
, CONCENTRATION,
DELIVERY RATES, AND VOLUME ARE ALL
SUITABLE
FOR CONCURRENT AND
PIGGYBACK
DELIVERY MODES. SEE
SECTION 8 FOR INFORMATION ON
CONCURRENT FLOW, SEE SECTION 9
FOR DELIVERY RATE RANGES.
NOTE: In concurrent mode with certain alarm conditions, one
line may stop while the other line may continue to deliver.
1 With pumping started on line A, to program a
Concurrent delivery rate, select
[B].
The Delivery Mode field is
highlighted.
NOTE: If an entry was previously
programmed, RATE is highlighted. If
no previous entries, MODE is
highlighted.
If field reads “Piggyback”,
select
[Change Mode] to
switch to “Concurrent”.
A
A
B
B
PUMPING STOPPED
Rate
mL/hr
Vol Inf
mL
Options/
Vol Inf
5.
5
0.
1
0
0
!
MedSurg
B
PROGRAM
Mode
Rate
VTBI
Duration
mL/hr
mL
hr : min
Program
Options
Change
Mode
Cancel/
Back
Change using Change Mode
Piggyback
0
0
00:00
MedSurg
Drug
List
sec04.fm Page 15 Wednesday, June 8, 2005 8:22 AM