Chapter 2 Service
The HP 53131A/132A Calibration Procedures
2-20 Assembly-Level Service Guide
2
CAL: TI QUIK? Calibration
1 Press any one of the arrow keys until
CAL: TI QUIK?
is displayed.
2Press Enter key.
A scrolling message with instructions appears. Follow the instruction by
performing the following procedure
3 Connect the HP 8130A Pulse Generator output to Channel 1 of the
Counter as shown in Figure 2-3.
Equipment
HP 8130A Pulse Generator (or equivalent)
HP 10503A BNC Cable
Figure 2-3. Calibration Setup for CAL: TI QUIK?
4 Set the HP 8130A as follows:
PERIOD: 100 ns
WIDTH: 50 ns
High: 0.50 V Low :
0.50 V
Input Mode: Normal
NOTE
Any external timebase connected to the Counter is ignored during this
calibration.
HP53131A/132A
Counter
Channel 1
HP 8130A
Pulse Generator
Output