Chapter 1 Performance Tests
Test 6: Peak Volts, Channel 1 (HP 53181A Only)
1-68 Assembly-Level Service Guide
1
Test 6: Peak Volts, Channel 1 (HP 53181A
Only)
This test verifies the Peak Volts accuracy specification of the HP 53181A
Frequency Counter.
Equipment
HP 3325B Synthesizer
HP 10100C 50
Feedthrough
HP 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2Press Other Meas key until
VOLT PEAKS 1
is displayed.
3 Set CHANNEL 1 and CHANNEL 2 inputs to:
50
/1M
, 50
X10 Attenuate, Off
100kHz Filter, Off

Table of Contents