Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)
Assembly-Level Service Guide 1-51
1
Test 2: Channel 1 Frequency Sensitivity
(HP 53181A Only)
This set of tests verifies frequency sensitivity specifications of the
HP 53181A Frequency Counter.
Equipment
HP 8663A Synthesized Signal Generator
HP 1250-0780 Type N-to-BNC Connector
HP 10100C 50
Feedthrough
HP 10503A BNC Cables (2)
Counter Setup
1Cycle the POWER key to preset the Counter.
2 Set CHANNEL 1 input conditions to:
Trigger/Sensitivity,
AUTO TRG: OFF
Trigger/Sensitivity,
LEVEL: 0.000 V
50
/1M
, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off