EasyManua.ls Logo

HP 654A - Frequency Adjustments

HP 654A
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model
654A
Section
V
5-42.
TURN-
ON.
543.
Connect
the
654A
to
a
variable
power
line
supply
(set
for
115V
or
230V
as
appropriate).
5-44.
Zero
set
the
meter (
Paragraph
3-7),
turn
the
instrument
on (
Paragraph
3-8),
and
allow
30
minutes
warm-up
time.
5-45.
POWER
SUPPLY
VOLTAGE
ADJUSTMENTS.
a.
Connect
a
dc
voltmeter
to
the
power
supply
positive
output (
Al
Pin
12).
b.
Adjust
Al
R14 (+
30V
Adjust)
for
31, +/-
0.2V.
c.
Connect
the
dc
voltmeter
to
the
power
supply
negative
output (
Al
Pin
13).
The
voltage
should
be -
26, +/-
0.5V;
if
not,
change
the
value
of
Al
R15*
to
obtain
the
required
voltage (
increasing
the
value
of
A1R15*
makes
the
power
supply
voltage
less
negative,
and
vice-versa).
d.
For
power
supply
troubleshooting,
refer
to
Para-
graph
5-78.
5-46.
POWER
SUPPLY
REGULATION
AND
RIPPLE
CHECK.
a.
Connect
the
dc
voltmeter
to
the
power
supply
negative
output (
Al
Pin
13);
switch
the
FREQUENCY
RANGE
switch
to
X100
and
note
the
voltmeter
indication.
b.
Vary
the
power
line
voltage
from
103.5V
to
126.5V (
207V
to
253V
for
a
230V
power
line);
the
de
voltmeter
indication
should
remain
within
+/-0.5V
of
the
reading
noted
in
step
a.
c.
Adjust
the
line
voltage
to
103.5V (
207V
for
a
230V
power
line)
and
disconnect
the
dc
voltmeter.
d.
Connect
an
oscilloscope
to
the
power
supply
positive
output (
Al
Pin 12)
and
short
out
the
tuner
capacitor
(
Cl)
by
clipping
a
lead
from
the
solder
lug (
green/white
lead)
on
the
tuner
frame
to
the
chassis.
The
oscilloscope
ripple
indication
should
be
less
than
15mV
peak-
to-
peak.
DO
NOT
MOVE
THE
TWO
WIRES
BE-
TWEEN
TUNER
AND
RANGE
SWITCH;
THEIR
POSITION
WILL
AFFECT
FREQ.
CAL.
e.
Connect
the
oscilloscope
to
the
power
supply
negative
output (
Al
Pin
13);
the
ripple
indication
should
be
less
than
15mV
peak-
to-
peak.
f.
Disconnect
the
clip
lead
from
the
tuner.
5-47.
FREQUENCY
CALIBRATION
PROCEDURES.
DO
NOT
MOVE
THE
TWO
WIRES
BE-
TWEEN
TUNER
AND
RANGE
SWITCH
AFTER
CALIBRATION
IS
STARTED;
THEIR
POSITION
WILL
AFFECT
FREQ.
CAL.
5-48.
The
frequency
calibration
set-up
is
shown
in
Figure
5-14 (
do
not
make
the
set-up
at this
time);
the
frequency
should
be
continuously
monitored
at
the
rear
panel
COUNTER
OUTPUT.
Table
5-2
lists
the
accuracy
required
at
each
check
frequency;
adjustments
for
each
range
are
listed
in
Table
54.
Components
located
on
the
FRE-
QUENCY
RANGE
switch (
S2)
are
identified
on
the
switch
drawing
of
Figure
7-2.
Table
5-4.
Frequency
Adjustments
Increasing
value
of
these
components
decreases
frequency
and
changes
A2TP2
voltage
as
shown.
Adjust
variable
capacitors
clockwise
to
increase
capacitance,
counter-
clockwise
to
decrease
capacitance.
FREQUENCY
RANGE
FREQUENCY
Increases
A2TP2
Voltage
Dial =
1
Decreases
A2TP2
Voltage
FREQUENCY
Increases
A2TP2
Voltage
Dial =
10
Decreases
A2TP2
Voltage
X10
X100
X1K
X1OK
X100K
X1M
S2R2*
- - -
- - -
.
- - -
S2R8*
S2R10*
- - -
- - -
---
---
S2R16*
- - -
S2C17*
S2C8,
S2C9*,
S2C13*,
S2C16*
S2C1*,
S2C18*
S2C10,
S2C11*
A2C1 ,
S2C12,
S2C15*
- - -
- - -
S2C2,
S2C3*
- - -
S2C4,
S2C5*
S2C6,
S2C7*
5-13

Related product manuals