EasyManua.ls Logo

HP 654A - 5-70. Troubleshooting the 654 A; 5-74. Front Panel Checks

HP 654A
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model
654A
Section
V
TEST
OSCILLATOR
hp654A
654A-8-1902
I
DB/STEP
ATTENUATOR
hp
355C
THERMAL
CONVERTER
hpl
10504
e
n:
DC
NULL
VOLTMETER
hp
4194
O O
r-
®
REFERENCE
I
0
SUPPLY
I
SEE
FIGURE
5-4
I
Figure
5-16.
Meter
Tracking
Adjustment
i.
Check
the
654A
meter
tracking
as
follows:
1)
Set
the
external
attenuator
to -
1dB
position
and
adjust
the
654A
AMPLITUDE
control
for
null
indication
on
the
dc
null
voltmeter;
the
654A
meter
should
indicate
0 +/-
0.05dBm.
2)
Set
the
external
attenuator
to
0
dB
position
and
adjust
the
654A
AMPLITUDE
control
for
null
indication
on
the
dc
null
voltmeter;
the
654A
meter
should
indicate -
1 ±
0.01
dBm.
3)
If
the
tolerances
in
steps
1)
and
2)
are
not
met
then
readjust
A3R7 (
Meter
Offset
Cal)
as in
step
c
and
repeat
steps
d
through
i
of
this
paragraph.
j.
The
meter
tracking
adjustment
is
completed
by
performing
Paragraph
5-60.
k.
For
troubleshooting
information
refer to
Para-
graph
5-94.
5-60.
AMPLITUDE
CONTROL
AND
METER
CALIBRATION.
NOTE
Perform
adjustments of Paragraph
5-59
before
making
the
following
adjustments.
a.
Connect
the
equipment
shown
in
Figure
5-1
using
the
50
ohm
termination.
b.
Set
the
654A
controls
as
follows:
FREQUENCY
dial
1
FREQUENCY
RANGE
X1K
OUTPUT
LEVEL
dBm +
10.0
IMPEDANCE
50
UNBAL
c.
Adjust
the
654A
AMPLITUDE
control
for
an
indication
on
the
ac
differential
voltmeter
of
.7071V
rms (+
10dBm);
adjust
A3R7 (
Meter
Offset
Cal)
for
a
654A
meter
indication
of
OdBm.
d.
Set
the
654A
AMPLITUDE
control
to
the
extreme
clockwise
position
and
adjust
A3R20 (
Amplitude
Cal)
for
an
ac
differential
voltmeter
indication
of
0.80V
rms (
approximately +
11.05dBm).
e.
For
troubleshooting
information
refer
to
Para-
graph
5-95.
5-61.
AMPLITUDE
ACCURACY
CALIBRATION.
a.
Perform
the
Amplitude
Accuracy
checks
of
Paragraphs
5-9
through
5-14
but
do
not
make
any
adjustments
except
as
outlined
below.
If
the +/-
1%
limits
are
met
for
all
impedances
do
not
perform
the
remaining
steps
of
this
Paragraph.
b.
If
the +/-
1%
limits
are
not
met
for
any
impedance:
1)
Verify
that
the
impedance
networks
on
the
Impedance
Selector
board (
A4)
have
the
correct
impedance
and
if
necessary
replace
any
faulty
resistors.
The
impedance
can
be
checked
by
measuring
the
open
circuit
voltage (
for
each
impedance)
which
should be
twice
the
terminated
voltage;
alternatively
set
the
attenuator
to -
80dB
position
and
measure
the
output
impedance
with
an
ohmmeter.
2)
Perform
procedure
of
Paragraph
5-60
except,
in
step
c (
of
5-60)
adjust
the
654A
AMPLITUDE
control
for
an
ac
differential
voltmeter
indication
slightly
higher
or
lower
than
0.707V
rms (
see
example
below)
so
as
to
correct
any
error
noted
in
step
a
of
this
Paragraph (
5-61):
the
offset
from
0.707V
rms
should be
as
small
as
possible,
but
in
any
case,
must
remain
within
the
limits
of
0.700
to
0.714V
rms.
5-17

Related product manuals