ADJUSTMENTS
-
--
5-22.
POWER CALIBRATION PROCEDURE (CHANGE
8)
(Cont'd)
6.
Set the power level to
+6
dBm, readjust
"0
MD
(A4R9)
to equal
+6
dBm measured
power.
7.
Iterate between steps
5
and
6
until both low and midpower ranges are calibrated and
no
readjustment is necessary.
8.
Set the power to
+
13 dBm as indicated on the plug-in display. Adjust
"0
HI" for a measured
power of
+
13
dBm.
9.
Step the
RF
power in
1
dB
intervals from
-2
to
+
13 dBm. The
RF
power at the 83522A
connector as read
on
the power meter should equal the indicated front panel
2.1
dBm.
If
necessary, readjust
"0
LO,"
"0
MID,"
and
"0
HI"
to calibrate power.