23
1. Turn on the equipment, and make sure there are not any objects on the lens cap.
2. If a message that says, “Initialize error” is displayed as shown in (Picture : Screen
showing an initialize error (1)), check to see there are any objects placed on the lens
cap or pinhole, and remove if there is any. Press “V” button and enter to measurement
screen. If the problem continues, inquire with authorized seller or manufacturer.
3. If a message that says, “Please remove the lens” is displayed as shown in (Picture :
Screen showing an initialize error (2)), remove the lens from the lens cap, and press
“V” button to enter to measurement screen.
4. If S, C and A are not 0, clean the pinhole (refer to Chapter 10.6 ‘How to Clean
Pinhole’), turn off the equipment and turn it back on. If the problem continues, inquire
with authorized seller or manufacturer.
6.3. Screen Saver
Screen saver will start automatically after a few minutes of non-operation. Screen saver
protects display unit by constantly changing the screen. Press 1~6 button or press MEM
button (save button) to go back to measurement screen. Do not turn off screen saver from
Set up screen without any apparent reason.