EasyManua.ls Logo

Hydthc CHC-1000S - Control Parameter Configuration

Hydthc CHC-1000S
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
宏宇达数控 & 弘宇达智能
地址:深圳市南山区中山园路 1001 E4-6CTCL 国际 E 城)
TEL0755-26625800 FAX0755-26729960 -17-
3.6.1 Control parameters
Press <Up> or <Down> to select the target menu item, press <ENTER> to enter the
[Control parameters] interface, the first page is as shown below:
the second page is as shown below:
The meaning of each parameter is shown in the following table:
Parameter Name
Definition
Measuring range
Set the maximum range that the system can measure. The default is
10mm. You can choose 10mm, 15mm, 20mm, 25mm range.
Dynamic
Response
Set the dynamic response of the system in mV/ms. The larger the
value, the faster the response. This parameter is valid only in sensor
mode (sensor controller only).
Parameter P
Set the PID parameter of the system. This parameter is valid only in
sensor mode (sensor controller only).
Parameter I
Set the PID parameter of the system. This parameter is valid only in
sensor mode (sensor controller only).
Parameter D
Set the PID parameter of the system. This parameter is valid only in
sensor mode (sensor controller only).
SetPBand
Setting the setpoint range (sensor controller only)

Related product manuals