EasyManua.ls Logo

iDig 2D Touch - Page 183

iDig 2D Touch
187 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration
183
Bad Bucket Calibration using the Automatic Procedure (cont.)
Repeat the calibration procedure:
Make sure to put the bucket tip/edge very accurately on the SAME spot each time.
Chose a solid reference spot so it won’t get moved in between the 8/10 points.
If the bucket sensor is on the dog bone, make sure to chose a high reference spot so you’ll be able to shoot points with the bucket more
closed than floating.
Chose the spots location in a reasonable distance to the cab, best results are 3-6 ft (1-2 m) away from the cab, referring to the
dipper/boom pivot point transferred vertically to the ground.
Bad Bucket Calibration using the Manual Procedure
Open/Close view not correct:
Make sure you selected the correct side for the bucket Sensor.
Re-check the bucket length:
Make sure to select the correct unit of length when entering the data.
Make sure to use the correct bucket pivot point to measure from.
Measure straight to the bucket tip/edge, do NOT measure as an angle.
Re-plumb the bucket:
Make sure that “Vertical” was selected.
Make sure that the plumb string line isn’t blocked by the quick link or bucket.
Make sure to choose the correct pivot point to attach the plumb line.
If you used the “Horizontal” option instead of the “Vertical” option, the bucket will be opened, NOT closed.