EasyManua.ls Logo

iDig 2D Touch - Page 25

iDig 2D Touch
187 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation Procedures Bucket Calibration
Manual Bucket Calibration
Press “Vertical” so that it is highlighted in blue. Select
which side the Bucket Sensor has been installed so that
it is highlighted in blue. NOTE sensor side refers to
which direction the plastic cap is facing. If its installed
on the left side but inside the coupler, the sensor is on
the Right Side. Press “Next.
The bucket will be named in your system as New
Bucket. You’ll need to go into the System Settings menu
to rename the bucket. Additionally, you’ll need to
perform the Floating Bucket procedure so that the
bucket on your screen matches the physical position of
the actual bucket. See next screen for these
instructions. Bucket calibration is done so press
“Done”!
25