EasyManua.ls Logo

IFM PI2897 - Finish calibration; Service functions; Reset all parameters to factory setting

IFM PI2897
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
a) Finish calibration:
Briefly press [Mode/Enter].
> [CAL] is displayed.
b) Change a 2nd point on the curve of measured values
Set a second defined reference pressure in the system.
Minimum distance between the calibration points CP1 and CP2 = 5 % of
the final value of the measuring range.
Press [Set] for 5 s.
> The pressure measured by the unit is displayed.
Press [Set] until the set reference pressure is indicated (measured
pressure = reference pressure) or the corresponding analogue signal is
provided on OUT2.
Maximum correction value = ± 2 % of the final value of the measuring
range.
Briefly press [Mode/Enter].
> [CP2] is displayed.
Briefly press [Mode/Enter].
> [CAL] is displayed, the process is finished.
9.5 Service functions
9.5.1 Read min/max values for the system pressure
Select [HI] or [LO] and briefly press [Set].
[HI] = maximum value, [LO] = minimum value.
Delete memory:
Select [HI] or [LO].
Press and hold [Set] until [----] is displayed.
Briefly press [Mode/Enter].
9.5.2 Reset all parameters to factory setting
Select [rES].
Press and hold [Set] until [----] is displayed.
Briefly press [Mode/Enter].
We recommend noting down your own settings before carrying out a reset
(→ 13 Factory setting).
9.6 Simulation function
9.6.1 Open menu level 3 (simulation)
Select [EF] and briefly press [Set] (= to open menu level 2).
Select [SIM] and briefly press [Set] (= to open menu level 3).
> [SEL] is displayed.

Related product manuals