Type of pressure: relative pressure
Order no. Measuring range
Permissible
overload
pressure
Bursting
pressure
bar psi bar psi bar psi
PI2893 -1...25 -14.4...362.7 100 1450 350 5075
PI2814 -1...16 -14.6...232 75 1085 250 3625
PI2894 -1...10 -14.5...145 50 725 150 2175
PI2815 -1...6 -14.5...87 30 435 100 1450
PI2895 -1...4 -14.5...58 30 435 100 1450
PI2896 -0.124...2.5 -1.8...36.27 20 290 50 725
PI2899 -1...1 -14.5...14.5 10 145 30 435
PI2897 -0.05...1 -0.73...14.5 10 145 30 435
mbar psi mbar psi mbar psi
PI2817 -100...1600 -1.46...23.2 15000 215 40000 580
mbar inH2O bar inH2O bar inH2O
PI2898 -12.4...250 -5.0..100.4 6 2400 30 12044
PI2889 -5...100 -2.00...40.16 4 1606 30 12044
Avoid static and dynamic overpressure exceeding the specified overload
pressure by taking appropriate measures.
The indicated bursting pressure must not be exceeded.
Even if the bursting pressure is exceeded only for a short time, the unit
may be destroyed. ATTENTION: Risk of injury!
The units are vacuum resistant.