Chapter: Requirements and Specifications
2-4 M10-16248-EN
Frame Location
The load frame must not be located against a wall or other object that interferes with air
ventilation around the frame. Proper air ventilation is required to dissipate the heat generated
from the frame base.
Floor Loading
The supporting floor must be checked for its weight loading and be able to support the
weight of the load frame and any other accessory parts that may have been purchased. The
floor must be free of vibrations from other activities taking place in or near the building.
For larger load frames, it is recommended that you consult your facility engineer, foundation
specialist, or a civil engineer to ensure that the load capacity of the floor can support at least
three times the weight of the load frame and its component parts. The higher load capacity is
necessary because large dynamic loads may occur when a specimen breaks and this
additional load may be absorbed into the floor.
Table Loading
The supporting table must be checked for its weight loading and must be able to support the
weight of the load frame with all its components and accessories, if the computer and
accessories are placed on the same table. There must be a sufficient margin of safety in the
table’s weight loading to ensure the stability of the system.
The table must be free of vibrations from other activities taking place in or near the building.
A table with leveling feet is recommended.
Power Supply
Ensure that the required power supply is available and complies with the requirements
detailed under “Power Requirements” on page 2-8.
Caution
Power fluctuations and surges can cause performance problems ranging from
intermittent electrical problems to severe damage.
Instron systems require “clean” and stable electrical power. Unless other arrangements are
made with Instron, you are responsible for providing clean electrical power. An in-line
power surge protector is recommended for all installations.
If you purchased an environmental chamber with your system, it requires a dedicated
electrical power supply that is separate from the power supply for the rest of the system.