Chapter: Maintenance
5-2 M10-16248-EN
Daily Maintenance Checks
Before operating the system each day, ensure:
• All cable connections are tight and secure.
• All grips, fixtures and accessories are free of dirt, damage and deformation.
• The load frame is level. If an adjustment is necessary, refer to “Level the Load
Frame” on page 4-1.
• Signal and power cables have adequate slack to prevent excessive strain on connectors.
• All cables are free of wear and chafing. Re-route the cables if necessary, and replace any
damaged cables.
• After turning on the system, make sure that power is adequately supplied to the
electronics.
Correct any problems before you operate the testing system. If you require assistance,
contact your local Instron Service department.
Periodic Inspections
Every six to twelve months, perform the following inspections:
• Visually inspect the machine for any loose fittings. Check the limit stops, cable
connections, and connections for any accessories that are attached to the machine.
Tighten any loose connections that you may find.
• Operate the crosshead through the full range of motion. It should move smoothly with
no unusual noise, or erratic motion. If it does not, refer to Table 5-1 on page 5-3 for
lubrication recommendations.
• Test the limit stops to ensure they are in working order. Refer to “Test Limit Stops” on
page 5-6.
• If you have any additional safety equipment added to the machine, test the equipment to
ensure it is in working order.
If you notice any problems resulting from these inspections, contact Instron’s Professional
Services department for immediate assistance.