EasyManuals Logo

Instron 5960 Series User Manual

Instron 5960 Series
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
Chapter: Installation
4-2 M10-16248-EN
5. When the machine is level, tighten the lock nut on each leveling foot.
Set the Input Voltage
The first step in the installation is to verify that the voltage and electrical plug are
compatible with the location’s power supply. The system voltage and electrical power plug
are set at the factory according to the power input voltage that was specified on your
purchase order. Use the following procedure only if the facility power source does not match
the frame voltage setting. This situation may arise if the system is moved to another location
with a different voltage rating from the factory setting on the frame.
Power Supply Compatibility
Verify the following before proceeding with installation:
The machine’s voltage is compatible with the main power supply to your facility.
The machine’s power cable can reach the electrical power supply with some slack in the
cable.
The plug is compatible with the electrical power outlet.
If your power source is not the voltage originally specified on your purchase order, follow
the instructions described in the following sections to change the machine’s voltage. Ensure
that you use the appropriate electrical plug when changing voltages.
Figure 4-1. Adjusting a Leveling Foot

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Instron 5960 Series and is the answer not in the manual?

Instron 5960 Series Specifications

General IconGeneral
BrandInstron
Model5960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals