Eclipse Additional Information Page 140
For optimum recordings, the EEG level should be low enough that a 20µV rejection threshold setting
could be applied and ensure that brief periods of increased EEG level would be rejected. However, for
non-optimum recording situations, a setting of 40µV may be needed. Such an action is not without conse-
quences however, and it is unlikely that near-threshold responses will be recorded in such adverse re-
cording conditions.

Table of Contents

Other manuals for Interacoustics Eclipse

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics Eclipse and is the answer not in the manual?

Interacoustics Eclipse Specifications

General IconGeneral
CategoryMedical Equipment
ManufacturerInteracoustics
OAEYes
ABRYes
StimulusClicks, Tone Bursts, Chirps
TransducerInsert earphones, Headphones
ConnectivityUSB
ApplicationsNewborn hearing screening
SoftwareEclipse Software Suite
Battery LifeNot applicable (mains powered)