Eclipse Additional Information Page 147
After attaching the electrodes to the patient, it is crucial to check the skin impedance. For best results, im-
pedance at each electrode should be as low as possible, preferably lower than 3k.
Press the Imp. button
and turn the dial fully
clockwise.
Turn the dial slowly
counter clockwise.
The LED for each
electrode will
change from red to
green as the imped-
ance is found for
each electrode.
Press the Imp. button to
exit the impedance mode
before testing.
Note If the impedance of one or more electrodes is too high wait for a minute or two as the gel on the elec-
trode has a tendency to improve the skin impedance over the first couple of minutes.
Return to recording mode by pressing down the IMP button once again (IMP light off).