Eclipse Additional Information Page 172
The checkmark will not be applied, even though appropriate S/N ratio has been met in the following situa-
tions:
If the S/N ratio has not been stable over a preset period of time.
If the DP level is below -10dB SPL. Such low level DPs do not carry the same diagnostic strength as DPs
of higher SPLs.
Select an individual point on the DPOAE Graph using the mouse to display the corresponding DP-Re-
sponse in the DP-Response graph.
Right click on the DPOAE Graph for the option to remove the checkmarks from the test points.
Change the display gain using the up and down arrows above the graph.