Eclipse Additional Information Page 188
8. A pre-set Screening Algorithm can be selected from the drop box.
9. Various Display Options of the Emission and noise floor can be selected. Selecting the Pass/Refer
Band option divides the emission and noise floor into five frequency bands. In the dropdown either the
OAE level or the Signal-to-noise-ratio can be selected to be displayed during recording.
10. Intensity can be set in 5 dB gradations between 50 and 90dB peSPL.
11. The Stop Criteria control the number of sweeps to be recorded. These can be set between 25 and
32000.
12. Wave Reproducibility allows the window in which reproducibility is calculated to be set between 2 milli-
seconds and 18 milliseconds. This can also be manipulated through the slide bar on the testing inter-
face.
13. The Default Rejection can be set between 25 and 55dB. This setting should be dictated by the amount
of background noise.

Table of Contents

Other manuals for Interacoustics Eclipse

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Interacoustics Eclipse and is the answer not in the manual?

Interacoustics Eclipse Specifications

General IconGeneral
CategoryMedical Equipment
ManufacturerInteracoustics
OAEYes
ABRYes
StimulusClicks, Tone Bursts, Chirps
TransducerInsert earphones, Headphones
ConnectivityUSB
ApplicationsNewborn hearing screening