Eclipse Additional Information Page 57
ABR Latency Norms
With the release of EP4.4 the latency norms for Click, CE-Chirp® LS and NB CE-Chirp® LS have been
aligned to make it more straight forward locating the wave V response. The figures below show how re-
sponse latencies for all three types of stimuli are aligned for the different stimuli levels. Note that age, rates
and filter settings can influence response latencies.
Latency Norms (mean) for the Click, CE-Chirp LS and NB CE-Chirp® LS (500-4 kHz) wave V ABR.
Latency norms for CE-Chirp® and NB CE-Chirp® are also present in the list if latency norms. The norm sets
are based on Elberling & Don, 2010; Kristensen & Elberling, 2012; Elberling et al, 2012a.
Click CE-Chirp® LS NB CE-Chirp® LS
Latency Norms for adults for Click, CE-Chirp® LS, NB CE-Chirp® LS (500-4 kHz) wave V ABR calculated with one
standard deviation (68.2%). The norm values are based on Kristensen & Elberling, 2012; Elberling et al. 2012a, Elberling