VisualEyes™ - Instructions for Use - EN Page 45
Figure 2.6-39: Improper detection of DataLink board (minilab 1008) as board#0 with System 2000 Comprehensive/Reclining chair
To solve this, change the board number by right clicking on the board in Instacal and click “Change
Board#...”
Configure instacal so the DataLink (miniLAB 1008) is Board#1, and the chair (USB-231 / PCI-DAS6025) is
Board#0.
Figure 2.6-40: Correct detection of DataLink board (minilab 1008) with System 2000 Comprehensive/Reclining chair