EasyManua.ls Logo

Interacoustics VisualEyes 505 - Page 83

Interacoustics VisualEyes 505
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D-0134090-A – 2022/11
VisualEyes™ - Instructions for Use - EN Page 78
Figure 3.7-10: Impedance Test screen.
When the impedance test has been performed, a standard calibration can be performed by following the
procedure in section 3.7.1.
If the electrode traces for horizontal and vertical traces show drift, the traces can be manually centered using
the Center Traces button. If the patient’s eye movements are not shown in the electrode traces, then the
Amplifier Gain can be adjusted below the traces to magnify the patient’s electrode signals (
Figure 3.7-11).

Table of Contents

Related product manuals