Setting Mode Menu Parameters
1. Press the PRINT/ACCUMULATE and ZERO buttons simultaneously. The display
will show the message ".". Press the PRINT/ACCUMULATE button. The
flashing display shows the current setting. Press the ZERO or the LOCAL/TOTAL
buttons to toggle between Auto, On and Off. When "Auto" is selected (default
setting), the backlight will automatically light up when low level light conditions are
detected. When the desired setting is displayed, press the PRINT/ ACCUMULATE
button to save.
2. The display will show "". Press the PRINT/ACCUMULATE button. The flashing
digit shows the current setting. When the weight displayed is equal to or greater
than the set point, the corresponding icon is displayed on the LCD screen. During
normal weighing mode, the S1 icon on the display will light up when the weight is
greater than or equal to Set Point 1. Use the LOCAL/TOTAL button to advance the
number and the ZERO button to move the flashing number to the left. When the
desired set point number is displayed, press the PRINT/ACCUMULATE button.
3. The display will show "". Press the PRINT/ACCUMULATE button. The flashing
digit shows the current setting. During normal weighing mode, the S2 icon on the
display will light up when the weight is greater than or equal to set point 2. Use the
LOCAL/TOTAL button to advance the number and the ZERO button to move the
flashing number the left. When the desired set point number is displayed, press the
PRINT/ACCUMULATE button.
4. The display will show "". Press the PRINT/ACCUMULATE button. The display
will show the current version of firmware loaded on the scale. Press the
PRINT/ACCUMULATE button.
5. The display will show the message ".". Press the PRINT/ACCUMULATE button.
The flashing digit shows the current setting. The number reflects the number of
readings that be averaged together before the reading is sent to the display. Higher
values will result in a more stable reading, but will take longer to process the final
value. The scale updates at 4Hz, so an Average Rate of "8" equates to 2 seconds of
averaging. Enter a "1" to disable averaging. Use the Lb/Kg button to the advance
the number and the ZERO button to move the flashing number to the left. When the
desired average rate setting is displayed, press the Mode button.
6. The display will show ".". Press the Mode button. The flashing digit reflects the
current Average Threshold setting. The setting enables dynamic averaging, which
can improve the settling time of a large Average Rate. If the scale detects a large
weight change, it will temporarily suspend averaging, jump to the new weight, and
resume averaging. Enter a value between 1 and 10000 to set the threshold (in
display divisions) at which the dynamic averaging activates. Enter "0" to disable
dynamic averaging. When the display shows the desired setting, press the
PRINT/ACCUMULATE button.