Invacare®TDX®SP2Series
2.1.3RepairandServiceInformation—Providersand/or
QualiedTechnicians
DANGER!
RiskofInjury,DamageorDeath
Impropersetup,service,adjustmentor
programmingmaycauseinjury,damageordeath.
–QualiedtechnicianMUSTsetup,serviceand
programthewheelchair.
–DONOTallownon-qualiedindividualsto
performanyworkoradjustmentsonthe
wheelchair.
–DONOTsetuporservicethewheelchairwhile
occupiedexceptforprogrammingorunless
otherwisenoted.
–TurnoffpowerBEFOREadjustingorservicing
thewheelchair.Notethatsomesafetyfeatures
willbedisabled.
–Ensureallhardwareissecurelytightenedafter
setup,serviceoradjustments.
–Warrantyisvoidifnon-qualiedindividuals
performanyworkonthisproduct.
DANGER!
RiskofDeath,SeriousInjury,orDamage
Corrodedelectricalcomponentsduetowater
and/orliquidexposure,orincontinentuserscan
resultindeath,seriousinjury,ordamage.
–Minimizeexposureofelectricalcomponentsto
waterand/orliquids.Electricalcomponents
damagedbycorrosionMUSTbereplaced
immediately.
–Wheelchairsthatareusedbyincontinent
usersand/orarefrequentlyexposedto
water/liquidsmayrequirereplacementof
electricalcomponentsmorefrequently.
WARNING!
–DONOTovertightenhardwareattachingtothe
frame.Thiscouldcausedamagetotheframe
tubing.
–Replacegas-lockingcylinderseverytwoyears
orifperformanceissuesareencountered.
Performanceissuesincludeforwardtippingand
veering.
WARNING!
RiskofInjury,DamageorDeath
Improperinstallationorservicemayresultin
injury,damageordeath.
–Transportreadypackagesarenotretrottable
toexistingmodelsandarenoteldserviceable.
–DONOTovertightenhardware.
14
60101831-A

Table of Contents