Invacare®TDX®SP2Series
7.2.15CleaningBatteryTerminals
WARNING!
RiskofInjury
Exposuretobatteryacidmayresultininjury.
–Theuseofrubberglovesisrecommended
whenworkingwithbatteries.
–DONOTallowtheliquidinthebatteryto
comeincontactwithskin,clothesorother
possessions.Itisaformofacidandharmfulor
damagingburnsmayresult.Shouldtheliquid
touchyourskin,washtheareaIMMEDIATELY
andthoroughlywithcoolwater .Inserious
casesorifeyecontactismade,seekmedical
attentionIMMEDIATELY .
–DONOTinstall/reinstallabatterywithacracked
orotherwisedamagedcase.
1.Examinebatteryterminalsforcorrosion.
2.Verifytheplasticcapsareinplaceoverbatterycell
holes.
3.Cleanterminalsbyusingabatterycleaningtool,wire
brush,ormediumgradesandpaper.
Uponcompletion,areasshouldbeshiny,not
dull.
4.Carefullydustoffallmetalparticles.
5660101831-A