Calibration Commands
Copyright © Itech Electronic Co., Ltd. 52
Chapter14 Calibration Commands
CALibrate:SECure:[STATe]
This command is used to set the power protection mode to be valid or invalid
during calibration.
Command syntax
CALibrate:SECure:[STATe] {<ON|OFF>,[<password>]}
Parameters
0|1|ON|OFF,
Example
CAL:SEC 1; CAL:SEC OFF
Query syntax
CALibrate:SECure:STATe?
Parameters
None
CALibrate:SAVe
This command is used to save the calibration coefficients in non-volatile
memory.
Command syntax
CALibrate:SAVe
Parameters
None
CALibrate:VOLTage:LEVel
The command is used to specify the voltage calibration point. P1, P2, P3, P4
calibration points must be sequentially calibrated.
Command syntax
CALibrate:VOLTage:LEVel <point>
Parameters
P1|P2|P3|P4