EasyManuals Logo

ITech IT7300 Series Programming Guide

ITech IT7300 Series
73 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
Calibration Commands
Copyright © Itech Electronic Co., Ltd. 54
Parameters
P1|P2
CALibrate:POWer [:DATa] {<numeric value>}
This command returns the actual output power value of the current calibration
point of the power supply.
Command syntax
CALibrate:POWer [:DATa] <NRf>
Parameters
<NRf>
Example
CAL:POW 80W
CALibrate:STRing
This command is used to set the calibration information during calibration.
Command syntax
CALibrate:STRing <Parameters>
Parameters
Support inputting a maximum length of 24 letters of string, which is recorded
the user calibration information. Such as the calibration time, frequency and so
on.
Example
CAL:STR "2005-1-9 20:12"
CALibrate:STRing?
This command is used to check the current calibration information.
Query syntax
CALibrate:STRing?
Returns
Calibration information saved in the power supply.

Table of Contents

Other manuals for ITech IT7300 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT7300 Series and is the answer not in the manual?

ITech IT7300 Series Specifications

General IconGeneral
BrandITech
ModelIT7300 Series
CategoryPower Supply
LanguageEnglish

Related product manuals