1. The Intelis Gas Meter records the Test Mode consumption in the Test Mode index
register.
2. The Intelis Gas Meter continues to output an optical pulse.
Note: While the Intelis Gas Meter is in Test Mode, the LCD refresh rate is more frequent
than when the meter is operating in Normal Mode.
3. The Intelis Gas Meter continues to update the normal volume registers.
Exiting Test Mode
The Intelis Gas Meter exits Test Mode after it receives a command from the FDM mobile
client, after the proving cable is removed from the meter, or when the Test Mode timeout is
reached. The following operations occur when the Intelis Gas Meter exits Test Mode.
1. The Intelis Gas Meter logs an Exiting Test Mode event.
2. The Intelis Gas Meter ends the optical pulse output.
3. The Intelis Gas Meter LCD returns to normal display mode.
Note: Check the LCD display to verify the Intelis Gas Meter is no longer displaying Menu
6: Test Mode.
Prover types
The Intelis Gas Meter compatible prover types are listed. See the procedure for your prover
type. For questions about your specific prover type, see the manufacturer's prover
documentation.
Prover type
Honeywell SNAP
Measurement Systems
Energy Economics (EEI)
Can-Tronics
Itron BPG SONICAL SNG6
Proving the Intelis Gas Meter using the Honeywell SNAP
pr
oving system
This section provides the information to set up an Intelis Gas Meter using a Honeywell
SNAP proving system.
Note: Proving the Intelis Gas Meter requires Honeywell SNAP prover software .NET X.Y.
Temper meters unpackaged in the proving area for 24 hours prior to testing. If tempering is
not possible, complete steps 1-9 of Tempering the meter by exercising on page 24. If the
meters are tempered, continue to Setting up the prover on page 25.
Proving the Intelis Gas Meter
Intelis Gas Meter Installation Guide TDC-1782-002 23
Proprietary and Confidential