EasyManua.ls Logo

ITRON Intelis - Chapter 7 Intelis Gas Meter Index Replacement; Index Assembly Replacement

ITRON Intelis
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10. (Recommended) Rerun the meter on the prover station to verify the meter calibration. It is
expected that you see the results that match the following example:
If the accuracy results are not as expected, repeat the calibration steps.
Note: Prover station accuracy is estimated to be +/- 0.15-0.25%.
T
o seal an Intelis gas meter
1. From the Tools menu, select OWR Intelis Gas Riva.
2. Select Test Mode Operations.
3. Select Seal Meter.
A progress bar indicates that the meter is sealing.
Adjusting the Intelis Gas Meter Calibration
Intelis Gas Meter Installation Guide TDC-1782-002 58
Proprietary and Confidential

Table of Contents

Other manuals for ITRON Intelis

Related product manuals