17
© 2017 JTECH Medical MN022_M
Performing a Full Calibration
(For all Instruments Except the Inclinometers and Goniometer)
Follow these steps to perform a full calibraon:
1. Select “Sengs” from the main menu.
2. Select “Instrument” from the Sengs menu.
3. Select the Instrument, for which you wish to perform a full calibraon.
4. Highlight the “System” opon in the Instrument menu.
5. Move the Control Sck le or right unl you highlight “Calibraon.”
6. Press the Control Sck to begin calibraon.
7. The Commander Echo Console will now aempt to connect to the selected instrument. Press the “On”
buon for the corresponding instrument followed by pressing the Control Sck on the Console to establish
connecon.
8. Follow the prompt to Press the Control Sck to set the instrument zero.
9. Move the Control Sck up and down to select the desired weight to be used for calibraon.
10. Once the proper weight has been selected, and this weight is currently being applied to the instrument,
press the Control Sck to enter the calibraon point.
Performing a Full Calibration on the Inclinometers
Follow These Steps to Perform a Full Calibraon:
Note: The angles for a full calibraon on the inclinometers must be exact for the calibraon to be valid. It is
recommended to use a T-Square or other tools that have exact right angles.
1. Select “Sengs” from the main menu.
2. Select “Instrument” from the Sengs menu.
3. Select “Inclinometer.”
4. Select the device ID for the Instrument which you wish to perform a full calibraon.
5. Using the joysck, scroll to the “System” opon in the Instrument menu.
6. Move the joysck le or right unl you highlight “Cal.”
7. Press the joysck to begin calibraon.
8. The Commander Echo Console will now aempt to connect to the selected instrument.
9. Press the “on” buon for the corresponding instrument to establish connecon.
10. Set the inclinometer in the upright posion.
11. Press the joysck to save this posion.
12. Rotate the inclinometer 90 degrees to the le.
13. Press the joysck to save this posion.
14. Rotate the inclinometer 90 degrees more to the le, so it is 180 degrees from the starng posion.
15. Press the joysck to save this posion.
16. Rotate the inclinometer an addional 90 degrees to the le, so it is 270 degrees from the starng posion.
17. Press the joysck to save this posion.
Performing a Full Calibration on the Goniometer
Follow These Steps to Perform a Full Calibraon:
1. Follow Steps 1-9 as above, selecng the goniometer in Step 3
2. Fully close the arm of the goniometer.
3. Press the joysck to set the instrument zero.
4. Open the arm of the goniometer to 180°
5. Press the joysck to save the calibraon