© 2017 JTECH Medical MN022_M
62
®
EMC Guidance
WARNING: Medical Electrical Equipment needs special precauons regarding EMC and needs to be
installed and put into service according to the EMC informaon provided in this manual. Carefully
read the informaon contained in this secon.
WARNING: The use of portable and mobile RF equipment can aect the normal operaon of
medical electrical equipment.
WARNING: Making any modicaons or using any accessories not specically approved by JTECH
Medical may reduce immunity to electromagnec interference or increase electromagnec
emissions.
Guidance and Manufacturer’s Declaraon - Electromagnec Emissions
The Commander Echo System is intended for use in the electromagnec environment specied
below. The customer or user of the Commander Echo System should assure that it is used in such
an environment.
Emissions test Compliance Electromagnec environment - guidance
RF emissions
CISPR 11
Group 2
The Commander Echo System must emit
electromagnec energy in order to perform its intended
funcon. Nearby electronic equipment may be aected.
RF emissions
CISPR 11
Class B
The Commander Echo System is suitable for use in all
establishments, including domesc establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domesc purposes.
Harmonic emissions
IEC 61000-3-2
Class A and D
Voltage uctuaons/
icker emissions
IEC 61000-3-3
Complies
WARNING: The Commander Echo instruments and Console should not be used while stacked on, or
adjacent to, other electrical or medical electrical equipment. If stacked or adjacent use is necessary,
all electrical equipment should be observed to verify normal operaon.