EasyManua.ls Logo

JTECH Medical COMMANDER ECHO - Page 63

JTECH Medical COMMANDER ECHO
69 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61
© 2017 JTECH Medical MN022_M
Technical Specications
Specified Power Supply (Battery Charger) (PW008*)
Used with the Daisy Chain for charging your Echo devices.
*Power supply connecon may vary according to the country shipped to
Operang Temperature, Humidity, and
Atmospheric Pressure
10 %
90 %
50 kPa
106 kPa
Transportaon and Storage Condions
10 %
100 %
50 kPa
106 kPa
Input Voltage
100-240 VAC
Input Frequency 50-60 Hz
Input Current 0.6 A
Enclosure Leakage Current 100 uA Normal Condion, 500 uA Single Fault Condion
Maximum Output Voltage/Current 5 V / 3 A
Maximum Output Power 15 W
Type of Protecon against electric shock
Class II
Manufacturer, Model JTECH Medical PN: PW008 [Globtek, Inc. WR9QA3000CCP-N-MED]
Protecon against harmful ingress of
water
IPX0 - ordinary equipment
Degree of Safety of Applicaon in the
presence of ammable mixtures
This device is not suitable for use in the presence of ammable anesthec
mixture with air, or with oxygen or nitrous oxide
Symbols
I
P
X
0
Dimensions 8.4 cm x 5.6 cm x 3.8 cm (3.3 in x 2.2 in x 1.5 in)
Weight 0.21 kg (0.46 lb)

Table of Contents