Message # Message Descripon
Message 11: Bad Time-Stamp, Secondary
Device
This message is displayed during tesng mode if the Console
receives an out-of-order data packet from the secondary
instrument. This ensures data integrity is maintained.
Message 12: Bad Sample Number, Primary
Device
This message is displayed during tesng mode if the Console
receives an out-of-order sample value from the primary
instrument, or in cases of RF interference. This ensures data
integrity is maintained.
Message 13: Bad Sample Number,
Secondary Device
This message is displayed during tesng mode if the Console
receives an out-of-order sample value from the secondary
instrument, or in cases of RF interference. This ensures data
integrity is maintained.
Message 14: Must Connect To Primary First This message is displayed if the user aempts to connect to a
secondary instrument (Secondary Inclinometer) without rst
connecng to a primary instrument for tesng.
Message 15: No Registered Instruments This message is displayed if the user aempts to connect to an
instrument when no instruments have been registered with
the Console.
Message 16: Test Data Corrupted, Test
Deleted
This message is displayed if corrupt test data is found while
the user is in review mode. The corrupt test will automacally
be deleted. This ensures data integrity is maintained.
Message 17: Primary Device Calibraon Is
Corrupted
This message is displayed if the Console receives corrupt
calibraon from the primary instrument before entering a
test. In some cases, if the user enters the sengs menu and
restores user calibraon then the problem will be xed. This
will work only if factory calibraon has not been corrupted.
If factory calibraon has been corrupted then the calibraon
values will have to be restored through a producon ulity or
through recalibraon.
Message 18: Secondary Device Calibraon
Is Corrupted
This message is displayed if the Console receives corrupt
calibraon from the secondary instrument before entering a
test. In some cases, if the user enters the sengs menu and
restores user calibraon then the problem will be xed.
This will work only if factory calibraon has not been
corrupted. If factory calibraon has been corrupted then
the calibraon values will have to be restored through a
producon ulity or through recalibraon.
Message 19: Zero Out Of Range This message is displayed if the Console receives an
instrument zero that is +/- 1 pounds from the factory zero
using factory calibraon informaon. This does not prohibit
the user from tesng.
Message 20: Calibraon Out Of Range This message is displayed if the Console receives instrument
calibraon data that is +/- 10% of the factory calibraon
informaon. This does not prohibit the user from tesng.
Message 21: No Free Instrument Slots This message appears if trying to register a new instrument
when 20 instruments have already been registered.