Message # Message Descripon
Message 11: Bad Time-Stamp, Secondary
Device
This message is displayed during tesng mode if the Console
receives an out-of-order data packet from the secondary
instrument. This ensures data integrity is maintained.
Message 12: Bad Sample Number, Primary
Device
This message is displayed during tesng mode if the Console
receives an out-of-order sample value from the primary
instrument, or in cases of RF interference. This ensures data
integrity is maintained.
Message 13: Bad Sample Number,
Secondary Device
This message is displayed during tesng mode if the Console
receives an out-of-order sample value from the secondary
instrument, or in cases of RF interference. This ensures data
integrity is maintained.
Message 14: Must Connect To Primary First This message is displayed if the user aempts to connect to a
secondary instrument (Secondary Inclinometer) without rst
connecng to a primary instrument for tesng.
Message 15: No Registered Instruments This message is displayed if the user aempts to connect to an
instrument when no instruments have been registered with
the Console.
Message 16: Test Data Corrupted, Test
Deleted
This message is displayed if corrupt test data is found while
the user is in review mode. The corrupt test will automacally
be deleted. This ensures data integrity is maintained.
Message 17: Primary Device Calibraon Is
Corrupted
This message is displayed if the Console receives corrupt
calibraon from the primary instrument before entering a
test. In some cases, if the user enters the sengs menu and
restores user calibraon then the problem will be xed. This
will work only if factory calibraon has not been corrupted.
If factory calibraon has been corrupted then the calibraon
values will have to be restored through a producon ulity or
through recalibraon.
Message 18: Secondary Device Calibraon
Is Corrupted
This message is displayed if the Console receives corrupt
calibraon from the secondary instrument before entering a
test. In some cases, if the user enters the sengs menu and
restores user calibraon then the problem will be xed.
This will work only if factory calibraon has not been
corrupted. If factory calibraon has been corrupted then
the calibraon values will have to be restored through a
producon ulity or through recalibraon.
Message 19: Zero Out Of Range This message is displayed if the Console receives an
instrument zero that is +/- 1 pounds from the factory zero
using factory calibraon informaon. This does not prohibit
the user from tesng.
Message 20: Calibraon Out Of Range This message is displayed if the Console receives instrument
calibraon data that is +/- 10% of the factory calibraon
informaon. This does not prohibit the user from tesng.
Message 21: No Free Instrument Slots This message appears if trying to register a new instrument
when 20 instruments have already been registered.