Pump set, order No. 9970140
Installation on the circulator
Push the end of the tubing on the "OUT" side on to the port
on the pump.
Secure against slipping using the clamp.
Push the end of the tubing on the "IN" side on to the holding
device on the pump.
Secure the pump housing to the base of the circulator using
the two screws.
Attach the circulator to the bath clamp.
The total immersion depth will be reduced due to the pump
set.
Connect an external system (also applies to bracket)
Remove the union nuts and sealing plates from the pump
connectors.
The hose connectors can be used for tubing with M16x1
connections in this state.
Or:
Secure barbed fittings to the union nuts.
Connect tubing and secure them with tube clamps to prevent
them sliding off.
Connect the tubing for the supply and return to the pump
connectors and the external consumer and secure them with
tube clamps.
Switch the pump function to external circulation.
Cooling coil, order No. 9970100
A cooling coil is required for working at around ambient temperature
(20 °C) A cooling water flow rate of 45 ml/min is generally sufficient
to compensate for the intrinsic temperature.
The cooling water temperature should be at least 5 °C lower than
the working temperature.