4 Functions and operation
Configuration level
You can access the configuration level from the parameter level by
pressing the q button (for at least 2 seconds).
The following functions can be programmed at this level:
-Probe type C111
-Linearization C112
-Number of decimal places C113
-Unit of measurement C002
-Supply frequency and temperature compensation C114
-Fixed value for temperature compensation C115
-Measurement display C116
-Function of the analog output C211
-Signal on probe break/short-circuit C212
-Function of the logic output 1 C221
-Signal on probe break/short-circuit C222
-Lower switching differential (hysteresis) C223
-Upper switching differential (hysteresis) C224
-Function of the logic output 2 C231
-Signal on probe break/short-circuit C232
-Lower switching differential (hysteresis) C233
-Upper switching differential (hysteresis) C234
H
The parameters can only be modified when the parameter
OFFS is active in the display and the configuration level has
been started. If another parameter is active, you can only
read the current setting.