TECHNICAL DESCRIPTION flowIQ® 3100
42 Kamstrup A/S • Technical Description • 5512-1242_B1_GB • 05.2015
11.2 Application environment
Ambient temperature: 2 °C…55 °C
Storage temperature: -25 °C…60 °C
Protection class: IP65
11.3 Lifetime
16 years, with a single AA battery
Battery change is possible
11.4 Connection of Pulse Adapter
From the factory the meter is configured not to transmit optical pulses otherwise, battery life will be
shortened. Therefore, having mounted the Pulse Adapter, the installer must set up the meter to transmit
optical pulses.
The procedure is as follows:
1. Unfasten the screw of the adapter ring to allow mounting of the unit on the meter
2. Dismount the two screws at the top and remove the cover
3. Push the button on the PCB, which is placed under the cover (see figure below)
Knap
LED
As soon as the button is activated, serial optical communication between Pulse Adapter and meter starts.
2
3
2
LED
Button
1