EasyManua.ls Logo

Kamstrup flowIQ 3100 - Measuring Principle - Flow Calculation Formulas

Kamstrup flowIQ 3100
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
flowIQ
®
3100 TECHNICAL DESCRIPTION
45
Kamstrup A/S • Technical Description • 5512-1242_B1_GB • 05.2015
12 Pulse Interface for flowIQ® 3100 / MULTICAL® 21
Pulse Interface, type 66-99-143, is used during calibration and verification in test stands with pulse
interface. The optical reading head is retained on flowIQ® 3100 by means of a transparent plastic holder
(Optical support, type 6561-331). Data from flowIQ® 3100 is read by the optical reading head, and in the
pulse interface unit converted into high-resolution volume pulses which can be registered by a pulse
receiver. When the optical reading head is removed, it takes 9 hours, before the meter returns to normal
mode.
Pulse Interface, type 66-99-143 , with Optical Support type 6561-331 on flowIQ® 3100
Supply: 3.5-30 VDC < 15 mA
Standby: < 0.2 mA
Pulse width: = 3.9ms
Frequency: Max frequency of 128 Hz
Meter size
Q3 [m
3
/h]:
Volume
[pulses/liter]
1.6 & 2.5 100
4.0 50
6.3 25
10 12.5
16 12.5
25 6